This tube furnace has a small appearance, high-quality resistance wire as heating element, double-layer shell structure and 30-segment program temperature control instrument, zero-crossing trigger, thyristor control, furnace chamber is made of alumina polycrystalline fiber material, furnace tube Both ends can be sealed with stainless steel flanges. The stainless steel flanges are equipped with gas nozzles, valves and pressure gauges. The vacuum degree can reach 10-3 Pa during vacuuming. It has a balanced temperature field, low surface temperature, fast rise and fall temperature, and energy saving. The advantages are ideal products for universities, scientific research institutes, and industrial and mining enterprises for high-temperature atmosphere sintering, atmosphere reduction, CVD experiments, and vacuum annealing.
Max. Temperature | 1200 degree |
---|---|
Working Temperature (Max.) | 1100 degree |
Furnace produce process:
Metal plate Cutting -Bending -Welding- Polishing – Powder coating- chamber board Carrving – Electrical module assembly – Whole machine assembly – Heating Test – Packing .
Item No.: | TF1200-I | TF1200-II |
---|---|---|
Rated Power: | 1.5KW | 1.5KW |
Heating Zone size (mm)(Dia. x L): | OD25 x 200mm | OD50 x 200mm |
Furnace tube size OD (mm): | 25×600 | 50×600 |
High purity quartz tube: 25/50*600mm (tube diameter is optional) The tube block is placed on each side. The tube block can block the heat radiation in the center of the furnace tube and must be placed in the correct position of the furnace before heating. ![]() |
||
Overall Dimension (WXDXH)mm: | 680X328X428 | 680X328X428 |
Rated Voltage: | 220V , 1 Phase. | 220V , 1 Phase. |
Heating Elements: | Domestic high quality resistance wire; Swedish Kanthal A1 imported resistance wire is optional; |
|
Control Mode : | 30-segment program temperature control: 1. The 30-segment program temperature control can only be adjusted by PID. 2. With over-temperature protection and broken couple protection, the heating circuit of the electric furnace is automatically cut off when over-temperature or broken couple; (When the temperature of the electric furnace exceeds 1200 degrees or the thermocouple is blown, the AC relay on the main circuit will be automatically disconnected, the main circuit will be disconnected, the ON light on the panel will go out, and the OFF light will be on, limited protection of the electric furnace) 3. With 485 communication interface (standard configuration when purchasing software) 4. With power-off protection function, that is, when the furnace is powered on after power-off, the program does not start from the initial temperature rise, but the furnace temperature starts to rise when the power is off. 5. The instrument has the function of temperature self-tuning. |
|
Chamber Materials: | 1. High-quality and high-purity alumina polycrystalline fiber curing furnace formed by vacuum suction filtration. 2. Using Japanese technology to form. 3. The spacing and pitch of the resistance wires in the furnace are all arranged according to the best thermal technology in Japan, and the temperature field is simulated by thermal software. |
|
Temperature Control Mode : | Intelligent PID | |
Control Accuracy: | ±1 ℃ | |
Thermocouple Type: | K Type | |
Trigger: | Zero crossing trigger | |
Thyristor: | German SEMIKRON 106/16E | |
Heating rate: | ≤30℃/ Min, suggest ≤15℃/ Min. | |
Temperature field curve: | The temperature field of TF1200 at a constant temperature of 900°C. ![]() |
|
Temperature calibration (optional): | Regular temperature calibration of the tube furnace can maintain the accuracy and uniformity of the furnace temperature. You can purchase a set of simple calibration tools to assist you in this task. A complete calibration kit includes standard thermocouples, temperature calibrators, and compensation wires. ![]() |
|
Valves and flanges: | ![]() |
|
Surface Temperature: | ≤45℃ | |
Over Temperature Protection: | Yes | |
Overcurrent Protection: | Yes | |
Broken Couple Protection: | Yes | |
Open cover protection system: | When the furnace cover is opened, the relay will automatically cut off the main power supply, effectively ensuring the safety of users. |
Vacuum system (optional):
1. Single-stage pump TW-1.5A :Generally used for rough vacuuming to exhaust the gas in the furnace tube.
2. Two-stage pump 2XZ-2 vacuum pump can pump the vacuum degree of the tube furnace tube to about 10Pa.
3. The VS-0.1 vacuum system can pump the vacuum in the tube furnace to 0.1Pa (and equipped with a resistance vacuum gauge display).
4. The VS high vacuum system (the system is composed of molecular and mechanical pumps) can pump the vacuum in the tube furnace to 10-3 Pa.
5. Constant pressure pump (this system can control the pressure in the pipe between -100.0KPa~100.0KPa).
6. UK (INFICON) capacitance vacuum gauge (measuring range 3.8×10-5 -1125 Torr, can detect various gases, including mixed gases, accurate measurement, good repeatability, optional anti-corrosion type PCG554)
7. You can purchase a cold vacuum trap from our company to avoid pollution to the vacuum pump。
Gas supply system (Optional):
1. Bubbler
The bubbler is particularly suitable for CVD experiments, allowing gas to enter the liquid in the tank through the gas inlet of the bubbler, bubbling the liquid (the inlet pipe nozzle is equipped with a bubbling device), thereby bringing liquid vapor into the reaction Cavity. The main body of this mixing tank is made of 316 stainless steel. The upper part of the tank is equipped with an air inlet and outlet and a mechanical pressure gauge (pressure range -0.1-0.15MPa). Two stainless steel stop valves are installed on the air inlet and outlet respectively. , Both joints adopt Φ6 ferrule joints (the joints can be connected to PTFE pipes or Dia6mm stainless steel pipes), and the pressure in the tank can withstand up to 0.15Mpa (the amount of liquid vapor increases in a positive pressure environment), At the same time, the tank can withstand a temperature of 250°C.
2. Mass flow meter gas supply system:2/3/5 or more gas channels optional;
3. Liquid evaporator
It is a system specially designed for the introduction of liquid in the CVD experiment in the laboratory. The liquid flow rate is controlled by a peristaltic pump. After the liquid is introduced into the gas mixing system by the peristaltic pump, it is heated by the heater in the system into steam. Then the hydrogen, nitrogen, or argon output from the gas supply system as the carrier gas is brought into the furnace tube. LEP can export a variety of liquids, such as: H2O, CH5OH, TiCl4r, SiHCl3, and Zn(C2H5)2, as well as a variety of organic compounds. This equipment is widely used in various CVD, catalytic reaction and activation experiments using non-corrosive liquid source steam.
Warranty period: One year warranty and lifetime technical support.
Note: 1. Consumable parts such as heating elements, thermocouple , quartz tubes, sample crucibles, etc. are not within the warranty period.
2. Damage caused by the use of corrosive gas and acid gas is not covered by the warranty.
Operating Notes:
1. The air pressure in the furnace tube cannot be higher than 0.02MPa;
2. Due to the high internal pressure of the gas cylinder, a pressure relief valve must be installed on the gas cylinder when gas is introduced into the furnace tube. In order to ensure safety, it is recommended that the pressure be lower than 0.02MPa;
3. When the temperature of the furnace body is higher than 1000℃, the furnace tube must not be in a vacuum state, and the air pressure in the furnace tube must be equal to the atmospheric pressure;
4. The gas flow into the furnace tube must be less than 200SCCM to avoid the impact of cold air flow on the hot furnace tube;
5. For the sample heating experiment, it is not recommended to close the exhaust valve and intake valve at the flange end of the furnace tube. If you need to close the gas valve to heat the sample, you must always pay attention to the indication of the pressure gauge. If the indication of the air pressure is greater than 0.02MPa, you must open the exhaust valve immediately to prevent accidents (such as furnace tube rupture, flange flying, etc.) .
The specification of the pressure reducing valve and the process of operation and ventilation during the use of the tube furnace:
1. The pressure in the furnace tube is maintained at 1 standard atmospheric pressure, and cannot be less than 1 standard atmospheric pressure, nor can it be greater than 0.02MPa. This method can be used to ensure that the pressure in the furnace tube is within this range.
2. The gas cylinder must be equipped with a pressure reducing valve. It is recommended to purchase a pressure reducing valve with a range of 0-0.6MPa. It is not possible to choose a pressure reducing valve with a large range (for example, 0-2.5MPa), this kind of pressure reducing valve is not accurate and easy to blow up the quartz tube.
3. It is appropriate for the pressure of the pressure reducing valve to open to 0-0.05MPa, and it should not exceed 0.05MPa.
4. When the furnace temperature exceeds 1000 degrees, vacuum sintering is not allowed in the furnace tube, and it is better to maintain a slight positive pressure atmosphere to protect the sintering.
5. The long-term use temperature of the quartz tube is recommended not to exceed 1100 degrees.
6. If you are worried that the air pressure is too high or the burned materials will emit gas, you can choose our company’s tube furnace explosion-proof system.
7. When the gas is ventilated, the gas outlet valve is adjusted to the maximum, and the gas flow can be adjusted by adjusting the gas inlet valve or flow meter.
8. It is not recommended to heat the furnace when the exhaust valve and intake valve at the flange end of the furnace tube are closed. If you need to close the gas valves at both ends to heat the sample (it is recommended to purchase our company’s tube furnace explosion-proof system), you need to always pay attention to the method. The indication of the blue pressure gauge should not allow the pressure indication to be greater than 0.02MPa. If the pressure indication is greater than 0.02MPa, the valve must be opened immediately to relieve the pressure to prevent the furnace tube from bursting or accidents.
Drainage and back-suction method: connect the air outlet pipe to the air inlet pipe of the conical flask short tube, connect the air outlet of the conical flask long tube to the pipe, and then put it into a beaker filled with water. Outdoors (if it is a toxic or dangerous gas), set the gas outlet valve to the maximum and adjust the gas flow through the inlet valve or flow meter. Because the water in the beaker has a certain depth, if the gas wants to come out, the gas is in the furnace. There must be a certain slight positive pressure inside. The insertion depth of the water pipe determines the pressure in the furnace tube. Since the air outlet valve is adjusted to the maximum, the pressure in the furnace body will not be too large (because the air outlet valve is larger than the air inlet valve), exceeding the water The pressure will come out. This method just achieves the effect of automatically adjusting the pressure balance in the furnace tube. You can observe whether the gas flows out and the speed of the outflow through the bubbling of the beaker. Make sure that there is gas out of the gas outlet, otherwise the water in the beaker will Pour into the Erlenmeyer flask. This Erlenmeyer flask acts as an anti-inhalation function to prevent water from directly entering the quartz tube.
The tube furnace uses Domestic high-quality resistance wire / imported resistance wire Sweden (kanthal) as heating element, adopts double-layer shell structure and electronic temperature control instrument, can carry out 30-segment temperature control, phase shift trigger, and thyristor control. The furnace adopts oxidation Aluminum polycrystalline fiber material has the advantages of balanced temperature field, low surface temperature, fast temperature rise and fall, and energy saving. It is an ideal product for universities, scientific research institutes, and industrial and mining enterprises for high-temperature atmosphere sintering, atmosphere reduction, CVD experiments, and vacuum annealing.
Detailed description:
Item: | Qty |
---|---|
Furnace Body: | 1 Set |
Quartz tube : | 1pc |
Stainless steel flange: | 1set |
Themocouple : | K type 1pc, Already installed |
Stainless steel hook: | 1pc |
Tube block | 1pc |
Alumina Crucible: | 1pc |
flange support: | 1pc |
High temperature gloves: | 1 pair |
COM to USB interface (Standard configuration when software control system is selected): | 1 pc |
software (Standard configuration when software control system is selected): | 1 pc |
User Manual : | 1PC |
OPTIONAL ACCESSORIES:
Gas Supply System: | Float flow meter air supply system | ![]() |
|
Proton flow meter gas supply system | ![]() |
||
Gas extraction system: | Vacuum Pump | TW-1.5A | ![]() |
2XZ-2 | ![]() |
||
VRD-4 | ![]() |
||
Pfeiffer molecular pump | ![]() |
||
Vacuum measurement: | Resistance vacuum gauge | ![]() |
|
Compound vacuum gauge | ![]() |
||
Bellows: | KF25 | ![]() |
|
KF40 | |||
Vacuum flapper valve: | KF25 | ![]() |
|
KF40 | |||
KF clamp: | KF25 | ![]() |
|
KF40 | |||
KF Tee joint: | KF25 | ![]() |
|
KF40 |